BAX® Q7: How to Run a Thermal Test
The video demonstrates how to perform a BAX® Q7 thermal test, which checks the functionality of the top heater and heated sample block.
Steps to Perform a Thermal Test:
- Cool Down the Cycler – Turn off the Bax Q7 cycler and wait 15 minutes before restarting.
- Start the Software – After 15 minutes, turn the cycler back on and open the Bax Q7 software.
- Run the Test –
- Click the Diagnostics tab (top left).
- Select Thermal Test (second to last option).
- Ensure the test lid box is checked, then click Run Test.
- Check for Completion – When the test is complete, blue text will indicate "Test Complete." A warning will also appear, stating the lid is hot.
- Cool Down Again – Turn off the instrument and wait another 15 minutes before proceeding with runs.
- Report Results (if requested by Hygiena) – Take a screenshot of the thermal test dialogue box and email it to diagnostic support with your case number.