PRO-Clean® Protein Residue Test: Instructional Video

This instructional video demonstrates the correct use of Hygiena®'s PRO-Clean protein residue detection tests.

  • If stored in cold conditions, allow the test to reach room temperature before use.
  • Remove the device from the foil pouch. Optionally, label it with location or time.
  • Twist and pull the swab from the tube.
  • Swab a 4-inch square area in a crisscross pattern, applying pressure to flex the wand and rotating for full coverage.
  • For irregular surfaces, ensure a consistent technique and sample a sufficient area.
  • Return the swab to the tube and break the snap valve by bending the bulb back and forth.
  • Squeeze the bulb twice to release liquid, then shake for 5-10 seconds.
  • Keep the tube upright during the reaction time.
  • After the reaction time (60 seconds), compare the liquid color to the chart on the label.
  • Store between 35°F - 77°F (2°C - 25°C). Detection levels vary with time: 0.80 µG protein (1 min), 0.50 µG (5 min), 0.20 µG (10 min).
  • For more details, contact Hygiena.